|
Indian Pediatr 2018;55:297-300 |
 |
Correlation Between
Blood Pressure Measurement by Non-invasive and Invasive Methods
in Critically-ill Children
|
B Vamshi Krishna, Suman Das and Sandip Sen
From Dr. BC Roy Post Graduate Institute of Pediatric
Sciences, Kolkata, West Bengal, India.
Correspondence to: Dr Suman Das, 44 Talpukur Road,
Deulpara, Naihati, North 24 Parganas,
West Bengal, Kolkata, India.
Email: [email protected]
Received: February 22, 2017;
Initial review: June 19, 2017;
Accepted: January 20, 2018.
Published online: February 09, 2018.
PII:S097475591600112
|
Objective: To determine the correlation of non-invasive blood
pressure obtained by auscultatory and oscillometric methods, with
invasive blood pressure in critically ill children. Methods: We
compared invasive with auscultatory and oscillometric blood pressures
using paired t-test, Pearson’s correlation coefficient and Bland-Altman
plot in 50 children (age 1-12 y) admitted in Pediatric intensive care
unit. Results: Systolic, diastolic, and mean arterial pressures
of invasive methods significantly correlated with auscultatory and
oscillometric methods (P<0.001). Auscultatory and oscillometric
measurements under-estimated systolic arterial pressures [mean (SD)
difference 5.4 (12.2) mmHg and 6.3 (14.0) mmHg, respectively; P<0.001]
and overestimated diastolic arterial pressures [-4.1 (5.8) mmHg and -3.6
(7.2) mmHg; P<0.001] compared to invasive blood pressure.
Conclusion: Mean arterial pressure obtained by NIBP measurement is
more closer than systolic or diastolic pressures, when compared with
invasive blood pressure measurement.
Keywords: Arterial catheter, Auscultatory, Oscillometry,
Sphygmomanometer.
|
A ccurate blood pressure (BP) values are required
for important diagnostic and therapeutic decisions in pediatric
intensive care units (PICU). Arterial cannulation with continuous
pressure transduction remains the reference standard for arterial BP.
The manual auscultatory BP (ABP) technique is uncommonly used in
critically ill children, except in emergencies or transports. Because of
the frequent need for repeated BP monitoring in PICU, oscillometric
non-invasive blood pressure (NIBP) measurements are more commonly used
[1,2]. Oscillometric blood pressure (OBP) monitoring offers the
advantage of automatic and direct measurement of mean arterial pressure
(MAP). In view of scarce data, validation studies are needed to compare
the precision of NIBP with invasive blood pressure (IBP) in critically
ill children[3]. We, therefore, determined the correlation and
reliability of NIBP obtained by auscultatory and oscillometric methods
compared to IBP.
Methods
We conducted the study at Dr BC Roy Post Graduate
Institute of Paediatric Sciences, Kolkata, after Institutional ethics
committee approval. Treating pediatricians took the decision to place an
intra-arterial catheter, without any influence from the investigators.
In the PICU, every child aged 1-12 years, and having a radial arterial
catheter for continuous IBP monitoring was selected, till we enrolled 50
patients for the study (consecutive sampling technique) after obtaining
written informed consent from guardians. Patients with contraindications
to cuff application/inflation (arm injuries or wounds, limb edema),
presence of arrhythmias, at high risk for intravascular thrombosis
(diabetic ketoacidosis, nephrotic syndrome and hypercoagulable states),
non-functional arterial catheter (defined as presence of overshooting or
undershooting phenomenon following rapid flush test), absent pulses and
having confounders of NIBP monitoring (chronic hypertension, obstructive
airway disease and obesity) were excluded.
The investigators filled the study proforma including
general details of children, history, general and systemic examination
findings, diagnosis, and treatment being received. A single investigator
measured BP in all participants to avoid inter-observer variability,
initially recording the NIBP followed by the IBP, to avoid the bias
introduced by the investigator being aware of IBP readings. We took 4
pairs of BP measurements from 50 patients, thereby collecting 200 pairs
of BP measurements (100 comparisons between IBP and ABP and 100 between
IBP and OBP).
For measuring IBP, the arterial catheter of proper
size (HMD KitKath, Germany) was inserted into the radial artery after
performing the Allen test and was connected to a disposable pressure
transducer (TruWaveTM, Edwards Lifesciences, Irvine, CA, USA) using
rigid pressure tubing of identical length. The transducer was connected
to the blood pressure module of the Phillips IntelliVue MP30 (Philips
MedizinSysteme, Germany) bedside monitor and to a heparinized saline (2
units/mL) syringe running at the rate of 3 mL/h, to prevent clotting of
the catheter. The pressure monitoring set had an Intraflo continuous
flush element pigtail that could be pulled to allow rapid flush of the
system. Before IBP measurements, the transducer position at the level of
the patient’s 4th intercostal space at the mid-axillary line, regulating
the pressure reading to zero, ensuring absence of kinking or air bubbles
in the tubing and transducer, flushing the tube and performing a "fast
flush" test to verify optimal damping and the presence of a normal
arterial waveform.
We measured all ABP in the brachial artery using a
well-calibrated aneroid sphygmomanometer (Smart Care, Indore, India)
with proper cuff sizes [4]. We chose an appropriately sized cuff
following a measurement of the arm circumference (at the midpoint
between the acromion and the humeral epicondyle). The investigator
positioned to see the dial of the manometer perpendicularly at eye
level. In case of persistence of the fifth Korotkoff tone, attenuation
of the fourth tone determined the diastolic arterial pressure (DAP). To
calculate the MAP from the values obtained by the sphygmomanometer, we
used the formula (SAP + 2DAP)/3 [systolic arterial pressure is
abbreviated as SAP]. We measured all the OBP in the brachial artery
using Phillips EasyCare cuffs (Philips Medical Systems, Andover, MA,
USA) of appropriate sizes [4] and connected with the NIBP pressure cable
into the BP module of the Phillips MP30 monitor. The NIBP was measured
in a different limb to the one with the arterial line. A 5-minute
interval separated one assessment from the other to avoid the
compression applied to the arm possibly affecting the measurements.
Statistical analysis: Data were entered into a
Microsoft Excel spread sheet and analyzed using SPSS version 20.0 [IBM
Corp., NY, USA]. The correlations of systolic, diastolic and mean
pressures of ABP and OBP versus IBP were investigated using
Pearson’s correlation coefficient, and agreement with Bland-Altman
analysis [5]. Paired t-test was used for comparisons between means. The
criterion for statistical significance was P<0.05.
Results
We screened 248 children; 198 were excluded (no
arterial line 153, arterial line not working 4, presence of arrhythmias
24, contra-indication to cuff application/inflation 1, and refusal of
consent 16). Finally we enrolled 50 (28 girls) children for the study.
The primary diagnosis of the study children were as
follows: 1 with acute gastro-enteritis and severe dehydration, 10 with
acute encephalitis syndrome, 4 with acute pancreatitis, 4 Congestive
cardiac failure (2 myocarditis, and 2 dilated cardiomyopathy), 1
disseminated tuberculosis, 2 near drowning, 3 empyema thoracis with
septicemia, 4 Guillain-Barre syndrome, 1 hanging, 1 head injury, 2
post-operative complications (perforated appendix), 5 pyogenic
meningitis, 3 pneumonia, 3 septicemia, 1 status asthmaticus, 4 status
epilepticus and 1 tubercular meningitis. Mean (SD) age of studied
children was 4.8 (3.2) years; 23 (46%) were 1-3 years, 11(22%) were 3-6
years, 12 (24%) were 6-9 years, and rest were >9 years. Fourteen
children (28%) received vaso-active drugs to treat hypotension; 4 each
received only dopamine or dobutamine, 3 received dopamine and
epinephrine, and rest 3 received dopamine and nor-epinephrine. None had
hypertension.
TABLE I Blood Pressure Differences Using IBP vs ABP
Pressures |
IBP |
ABP |
IBP-ABP |
P |
|
Mean (SD) |
Mean (SD) |
Mean (SD) |
value |
|
|
|
(95% CI) |
|
Systolic |
91.6 (24.3) |
86.2 (15) |
5.4 (12.2) |
<0.001 |
|
|
|
(3.0; 7.9) |
|
Diastolic |
51.2 (14.3) |
55.3 (11.7) |
-4.1 (5.8) |
<0.001 |
|
|
|
(-5.3; -3.0) |
|
Mean |
64.4 (17.3) |
65.6 (12.2) |
-1.2 (7.0) |
0.086 |
|
|
|
(-2.6; 0.18) |
|
IBP: invasive blood pressure; ABP: auscultatory blood
pressure. |
The differences between means of SAP, DAP and MAP of
IBP versus ABP and OBP, and results of paired t-test were
shown in Table I and II. We found a statistically
significant correlation between SAP, DAP, and MAP measurements of IBP
versus ABP, and OBP. Comparing IBP versus ABP, the Pearson’s
coefficients for SAP, DAP and MAP were 0.914 (P<0.001), 0.920 (P<0.001),
0.944 (P=0.0001). Comparing IBP versus OBP, the Pearson’s
coefficients for SAP, DAP and MAP were 0.908 (P<0.001), 0.866 (P<0.001),
0.916 (P<0.001). The Bland-Altman analysis (Table III)
showed wide limits of agreement between SAP, DAP and MAP of IBP
versus ABP and OBP.
TABLE II Differences Between SAP, DAP, MAP Using IBP vs OBP
Pressures |
IBP |
OBP |
IBP-OBP |
P-value |
|
Mean (SD) |
Mean (SD) |
Mean (SD) |
|
|
|
|
(95% C.I.) |
|
Systolic |
92.0 (25.0) |
85.6 (13.3) |
6.3 (14.0) |
<0.001 |
|
|
|
(3.6; 9.1) |
|
Diastolic |
50.6 (14.1) |
54.2 (10.5) |
-3.6 (7.3) |
<0.001 |
|
|
|
(-5.0; -2.1) |
|
Mean |
64.1 (17.3) |
64.5 (10.7) |
-0.5 (8.6) |
0.571 |
|
|
|
(-2.2; 1.2) |
|
IBP: invasive blood pressure; OBP: oscillometric blood
pressure. |
TABLE III Bland-Altman Analysis Showing Mean Differences (95% Limits of Agreement) Between IBP versus ABP and OBP
Differences |
Systolic |
Diastolic |
Mean |
IBP-ABP(95% |
5.4 |
-4.1 |
-1.2 |
limits of agreement) |
(29.3; -18.6) |
(7.2; -15.4) |
(12.6; -15) |
IBP-OBP(95% |
6.34 |
-3.58 |
-0.5 |
limits of agreement) |
(33.8; -21.1) |
(10.6; -17.8) |
(16.4; -17.4) |
IBP: invasive blood pressure; ABP: auscultatory blood
pressure; OBP: oscillometric blood pressure. |
Discussion
Both ABP and OBP significantly underestimated SAP and
overestimated DAP compared to IBP. The differences between mean MAP of
both ABP and OBP with IBP were small and not significant, but with wide
variability in individual differences. These differences depend on the
site of BP measurement as brachial IBP monitoring tends to be closer to
central BP than radial IBP. Inappropriateness of cuff sizes and rapid
deflation also contribute to discrepancies in auscultatory method [3,6].
However, these factors were minimized by following the proper method
strictly. The overestimation of diastolic pressure by non-invasive
methods may be due to cuff inflation. When cuff pressure exceeds venous
pressure, it occludes venous return, resulting in increased blood volume
in the arm distal to the cuff. This impairs diastolic run-off of blood
and elevates diastolic pressure [7]. The oscillometric methods are not
standardized and measuring algorithms differ among manufacturers and
devices, resulting in differences in OBP measurements [8,9].
The major limitation of the study is the comparison
of only non-invasive intermittent manual and automated techniques with
IBP, and inability to compare non-invasive automated continuous
techniques, due to unavailability of those devices in our institute. The
formula used for the calculation of MAP with the auscultatory method is
inaccurate in bradycardic or tachycardic patients, due to the length of
the systole changing with heart rate [10]. Unfortunately, no formula of
MAP adjusts for the heart rate. Analysis based on age- and sex-specific
normotensive, hypotensive and hypertensive criteria was not done.
Joffe, et al. [11] studied the difference
(NIBP–IBP) of SAP, DAP and MAP in 100 children. The mean difference was
usually small and significantly correlated. There was wide variability
in the differences between NIBP and IBP, and this varied even in the
same patient on the same day. Holt, et al. [12] compared IBP,
OBP, ABP and Doppler ultrasound measurements in 40 children. In the
normotensive range, there was no significant difference between systolic
pressures of NIBP and IBP, whereas mean and diastolic IBP measurements
significantly exceeded NIBP measurements [ DMAP
2.6, DDAP 5].
Outside the normotensive range, the automated readings were higher
during hypotension and lower during hypertension compared with the
invasive and Doppler ultrasound methods. Other researchers [13,14]
described the under-estimation of SAP and over-estimation of DAP by NIBP
techniques compared to IBP in adults. However, a study in neonates
testing the accuracy of NIBP device using SuperSTAT algorithm [15],
contradicted our results. They found that NIBP and IBP comparisons for
SAP, DAP and MAP met the 1992 US Association for the Advancement of
Medical Instrumentation accuracy standards, thereby concluding that NIBP
measurements taken with the new algorithm were accurate when compared to
IBP in neonates [15].
Contributors: SS: conceptualized the study; BVK:
collected data; SD: drafted the manuscript. All authors approved the
final manuscript.
Funding: None; Competing interest: None
stated.
What This Study Adds?
•
Non-invasive methods
significantly under-estimate systolic blood pressure and
over-estimate diastolic pressure; mean arterial pressure
obtained by non-invasive methods is more valid than either.
|
References
1. Parati G, Bilo G, Mancia G. Blood pressure
measurement in research and in clinical practice: recent evidence. Curr
Opin Nephrol Hypertens. 2004;13:343-57.
2. Tholl U, Forstner K, Anlauf M. Measuring blood
pressure: Pitfalls and recommendations. Nephrol Dial Transplant.
2004;19:766-70.
3. Chatterjee A, DePriest K, Blair R, Bowton D, Chin
R. Results of a survey of blood pressure monitoring by intensivists in
critically ill patients: A preliminary study. Crit Care Med.
2010;38:2335-8.
4. Bernstein D. Evaluation of the Cardiovascular
System. In: Kliegman RM, Stanton BF, Geme JW, Schor NF, editors.
Nelson Text Book of Pediatrics. 20th ed. Philadelphia: Elsevier; 2016.
p. 2164.
5. Bland JM, Altman D. Statistical methods for
assessing agreement between two methods of clinical measurement. Lancet.
1986;327:307-10.
6. Schroeder B, Barbeito A, Yosef SB, Mark JB.
Cardiovascular Monitoring. In: Miller RD, editor. Miller’s
Anesthesia. 8th ed. Philadelphia: Elsevier Saunders; 2015. p. 1347-9.
7. Hager H, Mandadi G, Pulley D, Eagon JC, Mascha E,
Nutter B, Kurz A. A comparison of noninvasive blood pressure measurement
on the wrist with invasive arterial blood pressure monitoring in
patients undergoing bariatric surgery. Obes Surg. 2009;19:717-24.
8. Pickering TG, Hall JE, Appel LJ, Falkner BE,
Graves J, Hill MN, et al. Recommendations for Blood Pressure
Measurement in Humans and Experimental Animals Part 1: Blood Pressure
Measurement in Humans: A Statement for Professionals from the
subcommittee of professional and Public Education of the American Heart
Association Council on High Blood Pressure Research. Hypertension.
2005;45:142-61.
9. Pickering TG. What will replace the mercury
sphygmomanometer? Blood Press Monit. 2003;8:23-5.
10. Razminia M, Trivedi A, Molnar J, Elbzour M,
Guerrero M, Salem Y, et al. Validation of a new formula for mean
arterial pressure calculation: The new formula is superior to the
standard formula. Catheter Cardiovasc Interv. 2004;63:419-25.
11. Joffe R, Duff J, Guerra GG, Pugh J, Joffe AR. The
accuracy of blood pressure measured by arterial line and non-invasive
cuff in critically ill children. Crit Care. 2016;20:1.
12. Holt TR, Withington DE, Mitchell E. Which
pressure to believe? A comparison of direct arterial with indirect blood
pressure measurement techniques in the pediatric intensive care unit.
Pediatr Crit Care Med. 2011;12:e391-4.
13. Ribezzo S, Spina E, Di Bartolomeo S, Sanson G.
Noninvasive techniques for blood pressure measurement are not a reliable
alternative to direct measurement: a randomized crossover trial in ICU.
Scientific World J. 2014;353628.
14. Davis RF. Clinical comparison of automated
auscultatory and oscillometric and catheter-transducer measurements of
arterial pressure. J Clin Monit. 1985;1:114-9.
15. Nelson RM, Stebor AD, Groh CM, Timoney PM,
Theobald KS, Friedman BA. Determination of accuracy in neonates for
non-invasive blood pressure device using an improved algorithm. Blood
Press Monit. 2002;7:123-9.
|
|
 |
|